MicroHR Motorized - A Compact, Versatile Czerny-Turner Spectrometer


The MicroHR is available in either a manual or motorized version.

Motorized Micro HR spectrometerAvailable as an imaging spectrograph or scanning monochromator, the new automated MicroHR allows users to make rapid and precise measurements, offering a degree of versatility not found in comparable focal length spectrometers. Equipped with an interchangeable dual grating turret, the automated MicroHR operates in a wavelength range that extends from 150 nm to 15 µm. As a spectrograph, it features a fixed or micrometer entrance slit with our unique Quick-Align CCD adapter. The addition of a side exit port increases versatility, allowing for dual detector operation; collect data from a CCD in the visible region then collect data in the IR using a single channel detector. The convenience of a USB interface allows users to scan and acquire data through a single computer connection. A rugged, high throughput optical system, the MicroHR is an ideal, affordable tool for a wide range of spectroscopic applications.

 

Features & Benefits

 

Features

Benefits

Compact and lightweight
Saves space
Easily integrated into optical systems
Czerny-Turner optical layout Optimal performance and minimal stray light
Automated, interchangeable dual grating turret Easy to switch wavelength range and upgrade in the field
Fast, high precision drive Allows for rapid, accurate measurements
Toroidal collimating mirror Provides best imaging quality
Pre-aligned CCD mount Allows for easy CCD installation without adjustment
Universal accessory mount Compatible with existing HORIBA Jobin Yvon accessories and C-mount
Works in all positions For versatility and portability

Optional side exit port

Dual detector mounting capability

Optional motorized internal filter wheel

Order sorting without complicated optics

USB 2.0 Interface Plug and Play computer interface

 

Specifications*

 

Focal length   140 mm
Entrance Aperture Ratio   F/3.88
Grating Size   32 mm x 32 mm

Grating Mount

 

Interchangeable dual grating turret

Image Magnification at Exit Slit

 

1.1

Scanning Range   0 - 1500 nm mechanical range
Drive Speed (increasing wavelength)   500 nm/s
Focal Plane   27 mm wide x 10 mm high
Spectral Dispersion   5.25 nm/mm at 400nm
Spectral Resolution   0.25 nm at 400nm
Wavelength Position Accuracy   +/- 0.25nm
Wavelength Repeatability (front exit)   +/- 0.04nm
Minimum Drive Step Size   0.025 nm at 400nm
Slits   Fixed or micrometer

Computer Interface

 

USB 2.0

Dimensions    Length 7" (178 mm)
Width

8.1" (206 mm) single exit port model
10.5" (267 mm) dual exit port model

Height 5.5" (140 mm)
Optical Axis Height 3.5" (89 mm)
Nominal Weight   10.0 lb (4.5 kg) single exit port model
11.0 lb (5.0 kg) dual exit port model

* All specifications with 1200 gr/mm grating, PMT with 12 µm slit width and 1mm height.

 

Schematic

Automated MicroHR Schematic

 
 
Back to Top