|
|
|
The Optics of Spectroscopy - Section 6: Entrance Optics
|
6.1 Choice
of Entrance Optics
While lenses are used in the examples that follow, front surface concave
mirrors coated for the spectral region of choice are preferred. A coating
such as aluminium is highly reflective from 170 nm to the near IR whereas
crown and flint glasses start losing transmission efficiency rapidly below
400 nm. "Achromatic Doublets" are routinely cemented with UV absorbing
resins and their anti reflective coatings often discriminate against
the UV below 425 nm. (This is due to the fact that such lenses are often
used in cameras where photographic film may be very UV sensitive).
If lenses must be used in the blue to UV , then choose uncoated quartz singlets
or air spaced doublets.

AS aperture stop
L1 lens 1
M1 mirror 1
M2 mirror 2
G1 grating 1
p object distance from lens L1
q image distance from lens L1
F focal length of lens L1
d the clear aperture of the lens (L1 in diagram)
The above diagram shows a typical monochromator system with one fixed exit
slit and one detector, however, all that follows is equally applicable to
a spectrograph.
6.1.1
Review of Basic Equations
Thin lens equation:
1/F = 1/p + 1/q (3-16)
Magnification (m):
(2-14)
For simplicity the diameter of an optic or that of its aperture stop (AS)
(assuming it is very close to the optic itself) is used to determine the
f/value. In which case Equations (2-4) and (2-5) simplify to:
f/valuein = p/d object f/value (6-1)
f/valueout = q/d image f/value
(6-2)
6.2
Establishing the Optical Axis of the Monochromator
System
6.2.1
Materials
6.2.2
Procedure
Assemble the above components so that the laser beam acts as the optical
axis which passes first through two pinhole apertures, followed by the monochromator,
and finally through the third pinhole aperture.
The external optics and source will eventually be placed on the optical axis
defined by the pinhole apertures and laser beam. Position the pinhole apertures
so that the lenses, etc. may be added without disturbing them.
Note: Reverse illumination may sometimes
be preferred where the laser passes first through the
exit slit and proceeds through all the optics until
it illuminates the light source itself.
Alignment of the components is an iterative process. The goal is for the
laser beam to pass through each slit center and to strike the center of each
optical element. The following steps achieve this:
-
If
a sine drive, then set the monochromator to zero
order.
-
Aim
the laser beam through the center of the entrance
slit.
-
Center
the beam on the first optic.
-
Center
the beam on the next optic, and so on until it passes
through the center of the exit slit.
-
If
the laser does not strike the center of the optic
following the grating, then rotate the grating until
it does. Many spectrometers are not accurately calibrated
at zero order, therefore, some offset is to be expected.
6.3
Illuminating a Spectrometer
If a light source such as a sample or a calibration lamp is to be focused
into the entrance slit of a spectrometer, then:
* Ensure that the first active optic is homogeneously illuminated. (Plane
mirrors are passive).
* Place a white screen between the entrance slit and the first active optic.
(In a CZ monochromator, the collimating mirror and in an aberration corrected
concave grating, the grating itself.) Check for "images", if there
is a uniform homogeneously illuminated area, all is well. If not, adjust
the entrance optics until there is.
6.4
Entrance Optics Examples
The majority of commercial spectrometers operate between f/3 and f/15, but
the diagrams that follow use drawings consistent with f/3 and all the calculations
assume f/6.
In the examples which follow, the lens (L1) used is a single thin lens of
100 mm focal length (for an object at infinity) and 60 mm in diameter.
The f/valueout of the entrance optics must be equal to the f/valuein of
the monochromator.
If necessary, an aperture stop should be used to adjust the diameter of the
entrance optics.
Remember when calculating the diameter of aperture stops, to slightly underfill
the spectrometer optics to prevent stray reflections inside the spectrometer
housing.
6.4.1
Aperture Matching a Small Source
Example 1 (Figure 23)
The emitting source is smaller in width than the width of the entrance slit
for a required bandpass.

1) Calculate the entrance slitwidth for appropriate bandpass (Equation
(3-9)). For this example, let the slitwidth be 0.25 mm.
2) Example Object: a fiber of 0.05 mm core diameter and NA of 0.25.
3) Object emits light at f/2 (NA = 0.25). Spectrometer = f/6.
4) Projected image size of fiber that would be accommodated by the system
(given by entrance slitwidth) = 0.25 mm.
Calculate magnification to fill entrance slit.
5) m = image size/object size = 0.25/0.05 = 5.0.
Therefore, q/p = 5, q = 5p.
6) Substituting into the lens Equation (3-16) gives p = 120 mm, and q = 600
mm.
7) To calculate d, light must be collected at f/2 and be projected at f/6
to perfectly fill the grating.
Therefore, p/d = 2, d = 120/2 = 60 mm.
Therefore, aperture stop = full diameter of L1.
Projection f/value = 600/60 = 10.
In other words, the grating of the monochromator, even though receiving light
collected at f/2, is underfilled by the projected cone at f/10. All the light
that could have been collected has been collected and no further improvement
is possible.
Example 2
If, however, the fiber emitted light at f/1, light collection could be further
improved by using a lens in the same configuration, but 120 mm in diameter.
This would, however, produce an output f/value of
600/120 = f/5
Because this exceeds the f/6 of the spectrometer, maximum system light collection
would be produced by a lens with diameter
d = q/(f/value) = 600/6 = 100 mm
thereby matching the light collection etendue to the limiting etendue of
the spectrometer.
The collection f/value is, therefore,
f/valuein = p/d = 120/100 = 1.2
Since etendue is proportional to the square of the (f/value)-1,
about 70% of the available emitted light would be collected at f/1.2. See
Section 3.
If the user had simply placed the fiber at the entrance slit with no entrance
optics, only 3% of the available light would have been collected. (Light
in this case was collected at the spectrometer's f/6 rather than the f/1.2
with etendue matching entrance optics).
6.4.2
Aperture Matching an Extended Source

The object width is equal
to or greater than the entrance slit width. See Figure
24.
The f/valueout of the entrance optics must be equal to the f/valuein of
the monochromator.
The object distance should be equal to the image distance (absolute magnification,
m, equals 1).
Aperture stops should be used to match etendue of the entrance optics to
the monochromator .
Because the object is larger than the slitwidth, it is the monochromator
etendue that will limit light collection.
In this case, image 1:1 at unit magnification.
l) Taking lens L1
So for F = 100 mm, p = 200 mm, q = 200 mm (2F).
2) f/value of the monochromator = q/d = p/d = 6.
3) Then
d = q/(f/value) = 200/6 = 33.3
Therefore, aperture stop = 33.33 mm to fill the diffraction grating perfectly.
6.4.3
Demagnifying a Source
In this case the f/value of the source is numerically larger than that of
the spectrometer. This is often seen with a telescope which may project at
f/30 but is to be monitored by a spectrometer at f/6. In this case etendue
matching is achieved by the demagnification of the source. See Figure 25.

1) Calculate the entrance slitwidth for the appropriate bandpass(Equation
(2-21)). Take, for example, 1.0 mm = final image size = entrance slit width.
2) Image projected by telescope = 5 mm and forms the object for the spectrometer.
m = 1/5 = 0.2,
then from Equation (3-16). Taking lens L1 with F = 100 mm (given),
p = 600 mm, q = 120 mm.
Calculate d knowing the monochromator f/value = 6.
q/d = 6, d = 120/6 = 20 mm.
The aperture stop will be 20 mm diameter.
Light is gathered at either the aperture of the projected image or 600/20
= f/30, whichever is numerically greater.
6.5
Use of Field Lenses
The concepts given in this section have not included the use of field lenses.
Extended sources often require each pupil in the train to be imaged onto
the next pupil downstream to prevent light loss due to overfilling the optics
(vignetting). See Section 2.8.
-
Used
when entrance slit height is large and the light
source is extended.
-
A
field lens images one pupil onto another. In Fig.
26, AS is imaged onto G1.
Field lenses ensure that for an extended source and finite slit height, all
light reaches the grating without vignetting. In Figures 26 and 27 the
height of the slit is in the plane of the paper.

6.6
Pinhole Camera Effect
When entrance optics are absent, it is possible for the entrance slit to
project an image of just about everything before the slit into the spectrometer.
This may include the lamp, the sample, rims of lenses, even distant windows. Section
3 describes how to correctly illuminate a spectrometer for highest
throughput. Following this procedure will eliminate the pinhole camera effect.
Multiple imaging may severely degrade exit image quality and throughput.
On the other hand, the pinhole camera effect is very useful in the VUV when
refractive lenses are not available and mirrors would be inefficient.
6.7
Spatial Filters
Aperture and field stops may be used to reduce or even eliminate structure
in a light source, and block the unwanted portions of the light (e.g., the
cladding around an optical fiber). In this capacity, aperture stops are called
spatial filters. See Figure 28.
The light source image is focused onto the plane of the spatial filter. which
then becomes the light source for the system.

|
|
|
|